SKU
HI96801

Digital Refractometer for Brix Analysis in Foods - HI96801

€253.00
In stock
The HI96801 is a rugged, portable digital refractometer for food products designed to report the sugar content of aqueous solutions as % Brix. The HI96801 report the results with an accuracy of ±0.2% Brix. The operation of the meter is simplified with only two buttons: one button is to calibrate with distilled or deionized water and the other to take a measurement. All readings are automatically compensated for temperature variations according to the ICUMSA Methods Book standard and displayed within a 1.5 second response time. The sealed flint glass prism and stainless steel well are easy to clean. Just wipe with a soft cloth in preparation for the next sample.

The HI96801 is a rugged, portable digital refractometer for food products designed to report the sugar content of aqueous solutions as % Brix. The HI96801 report the results with an accuracy of ±0.2% Brix. The operation of the meter is simplified with only two buttons: one button is to calibrate with distilled or deionized water and the other to take a measurement. All readings are automatically compensated for temperature variations according to the ICUMSA Methods Book standard and displayed within a 1.5 second response time. The sealed flint glass prism and stainless steel well are easy to clean. Just wipe with a soft cloth in preparation for the next sample.

Making a % Standard Sucrose Solution To make a Sucrose Solution, follow the procedure below:

  • Place container (such as a glass vial or dropper bottle that has a cover) on an analytical balance.
  • Tare the balance.
  • Weigh out X grams of high purity sucrose (CAS #: 57-50-1 directly into a beaker
  • Add distilled or deionized water to the beaker so the total weight of the solution is 100g.
  • Note: Solutions above 60% may need to be vigorously stirred or shaken and heated in a water bath to roughly 40°C (104°F). Remove solution when sucrose has dissolved and allow to cool before use.

 

Refractive Index

The HI96801 takes measurements based on a sample’s refractive index. Refractive index is a measurement of how light behaves as it passes through the sample. Depending on the sample’s composition, light will refract and reflect differently. By measuring this activity with a linear image sensor, the sample’s refractive index can be assessed and used to determine its physical properties such as concentration and density. In addition to the linear image sensor, the HI96801 uses an LED light, prism and lens to make measurement possible. Variations in temperature will affect the density of a solution based on the compound that is present. In digital refractometry the use of temperature compensation is necessary for accurate results. The HI96801 contains a built-in temperature sensor and is programmed with temperature compensation algorithms in accordance with the ICUMSA Methods Book for a percent by weight sucrose solution.

 

Features at-a-glance:

 

One-point calibration

  • Calibrate with distilled or deionized water

Small sample size

  • Sample size can be as small as 2 metric drops (100 μl)

Automatic Temperature Compensation (ATC)

  • Samples automatically compensated for temperature variations

Quick, accurate results

  • Readings are displayed in approximately 1.5 seconds

Dual-level LCD

  • The dual-level LCD displays measurement and temperature readings simultaneously
  • Stainless steel sample well Easy to clean and corrosion-resistant
  • Battery indicator Battery percent level remaining at startup and low battery indicator
  • Automatic shut-off To conserve battery life the meter shuts off automatically after three minutes of non-use
  • Stainless steel sample well Easy to clean and corrosion-resistant
  • IP65 water resistant protection Water resistant ABS plastic casing designed to perform under laboratory and field conditions.

 

Specifications

 

Specification Name Detail
SKU HI96801
Sugar Content Range 0 to 85% Brix
Sugar Content Resolution 0.1 % Brix
Sugar Content Accuracy ±0.2% Brix
Temperature Range 0.0 to 80.0°C (32.0 to 176.0°F)
Temperature Resolution 0.1°C (0.1°F)
Temperature Accuracy ±0.3°C (±0.5°F)
Automatic Shut-Off after three minutes of non-use
Battery Type / Life 9V / approximately 5000 readings
Enclosure Rating IP65
Dimensions 192 x 102 x 67 mm (7.6 x 4.01 x 2.6”)
Weight 420 g (14.8 oz.)
Refractometer Light Source yellow LED
Refractometer Temperature Compensation automatic between 10 and 40°C (50 to 104°F)
Minimum Sample Volume 100 μL (to cover prism totally)
Sample Cell stainless steel ring and flint glass prism
Measurement Time approximately 1.5 seconds
Ordering Information Supplied with battery and instruction manual.
Warranty 2 years